MPL1
Luminometer
|
|
The
MPL1 microplate luminometer is a high quality yet affordable instrument
with basic functionality. It incorporates all important functions
necessary to perform glow luminescence measurements and it is suitable
for all applications which require a wide dynamic range, but not
extremely high sensitivity.
|
|
Products
|
Request
brochure
|
Request
quotation
|
Home
|
|
Microplate Luminometers
MPL2
Orion L
Orion II
Single
Tube Luminometers
Luminometer Validation Tools
|

|
MPL1
applications include, but are not limited to:
Reporter Gene Assays
Quantification of Nucleic
Acids
Immunoasssays
ATP Assays
Cellular Luminescence
|
|
Technical
Data |
|
Detector
|
Photomultiplier
Tube with Bialkali Cathode, effective spectral sensitivity range
300-650 nm, operated in current measurement mode.
|
|
Sample
Format
|
96
well microplate, opaque in solid or strip format.
|
|
Sensitivity
|
ATP
Assay: 1 fmole (SL Assay; BioThema, Sweden).
|
|
Crosstalk
|
Less
than 3 x 10-5 (spillover from adjacent wells).
|
|
Dynamic
Range
|
6
decades.
|
|
Hardware
|
Micro-controller
for control of all basic instrument functions.
|
|
Measuring
Time
|
1s
per well. Cycle time for a complete microplate 120-200s.
|
|
Scan
Pattern
|
Random
Selectable single well, access by mouseclick.
|
|
Interface
|
Data
transfer via RS-232 cable. 9 pin male plug.
|
|
Dimensions
|
Instrument
- W: 400 mm, D: 490 mm, H: 139 mm
Transport Package - W: 500 mm, D: 610 mm, H: 260 mm
|
|
Weight
|
17
kg
|
|
Power
|
230
V 50 Hz; 115 V 60 Hz.
|
|
Power
Consumption
|
25
VA.
|
|
Storage
Temperature
|
0°C
- 40°C
|
|
Operating
Temperature
|
10°C
- 35°C recommended
|
|
Humidity
|
10%
- 80% (non-condensing)
|
|
|
PC software |
|
System
Requirements
|
IBM
Compatible PC, Pentium Processor; one
available serial port
|
|
Operating
System
|
Windows®
95, NT 4.0 or higher
|
|
Screen
Resolution/Colors
|
min.
640 x 480 x 16 colors
|
|
RAM
|
min.
16MB, min. 64MB for optimal use of graphs
|
|
Operating
System
|
Windows®
95, NT 4.0 or higher
|
|
Additional
Software
|
Microsoft®
ExcelTM; 97 or higher (optional)
|
|
Standard
Configuration
|
Protocol
Manager, Quick Measurement
|
|
Additionaly
Available
|
Slow
Kinetics. Please inquire for more protocols.
|
|
Optional
Software
|
MikroWin
2000
|
|
Windows
Features Provided
|
Full
OLE support, fully resizable grids,
non-contiguous areas selectable, multiple
document interface, resizable, zoomable
graphs, status bar, draggable toolbars,
|
|